Samples of Comparison for Visual Binary Testing of p-Chloroaniline as Impurity in Substance Chlorhexidine Digluconate
Анотація
For the visual detection of the impurity - p-chloroaniline (p-CA) at the level of its limiting content in the substance chlorhexidine digluconate (CH), it was suggested to use the screening methodology - binary testing using a single sample of comparison. Two indicator reactions were selected: the reaction in a solution of diazotized p-CA with N-naphthylethylenediamine (sample of comparison 1) and the chemisorption reaction of diazotized p-CA with polyurethane foam (sample of comparison 2). For establishing the concentration of p-CA in the samples of comparison a statistical approach that previously had used for estimating the limit of detection for visual test methods, was applied. Estimated value of the threshold concentration of p-CA less than the normilized level on the value which providing the risk of false-negative test result not more than 5 %. Accuracy of the visual binary testing of p-CA in the chlorhexidine digluconate substance was confirmed by spectrophotometric, chromatographic methods and diffuse reflection spectroscopy.
Як цитувати
E.A. Reshetnyak, A.Yu. Solokha, A.A. Khadzhikova, A.V. Panteleimonov. Samples of Comparison for Visual Binary Testing of p-Chloroaniline as Impurity in Substance Chlorhexidine Digluconate. Methods Objects Chem. Anal. 2017, 12(3), 123-129. https://doi.org/10.17721/moca.2017.123-129