Ways to improve the analytical characteristics of the atomizer "graphite" in atomic-absorption determination of some heavy metals
† Odessa I.I.Mechnikov National University, Dvoryanskaya Str., 2,Odessa, 65082 Email: A191047@i.ua
Abstract
This paper presents the model of the heating rate electrothermal graphite tube furnace, including graphite «filter-furnace» atomizer «GRAPHITE-2», as well as the appropriate operating parameters of the ET AAS de-termination Pb and Cd. It is shown that the use of a graphite «filter-furnace» atomizer provides a ~ 2.0-2.5 fold improve the sensitivity of determination Pb and Cd, increase the temperature at pyrolisis stage by ~ 100 – 150 0C, as well as completely eliminate interference effects including background absorbance by the presence in analyzed solutions of 0.5 g∙dm-3 NaCl, Na2SO4, CaCl2, MgSO4. As the result, simple and effective method for the direct ET AAS determination of ≥ 0.0015; 00001 mg·dm-3 Pb and Cd respec-tively, in natural waters are proposed. The relative standard deviation (Sr) does not exceed 7%
How to cite
A.N. ZACHARIA, A.S. ZHURAVLEV, V.V. KALINCHAK, A.S. CHERNENKO, A.N. CHEBOTAREV, M.V. ARABADZHI. Ways to improve the analytical characteristics of the atomizer "graphite" in atomic-absorption determination of some heavy metals. Methods Objects Chem. Anal. 2014, 9(1), 39-49. https://doi.org/10.17721/moca.2014.39-49